19
Set respective value and press key to confirm, next go to the home
screen.
The higher filter value, the longer the weighing takes.
13.2. Value Release
Enter this parameter to adjust rate of stabilisation of the measurement result.
Depending on the selected option, weighing time is either shorter or longer.
Procedure:
Enter <P2.rEAd / 2.2.APPr> submenu.
Press key, available values are displayed successively one by one:
F_P - fast and reliable, PrEc reliable, FASt fast.
Press key to confirm, next go to the home screen.
13.3. Ambient Conditions
Parameter relating to ambient and environmental conditions of the workstation.
Enter this parameter and set 'nStAb' value if the ambient conditions are
unfavourable (air drafts, vibrations).
Procedure:
Enter <P2.rEAd / 2.3.Enut> submenu.
Press key, parameter values are displayed successively one by
one: nStAb unstable, StAb stable.
Press key to confirm, next go to the home screen.
13.4. Autozero Function
'Autozero' function has been designed to enable automatic control and
correction of zero indication. This guarantees precise weighing results.
There are, however, some cases when this function can be a disturbing factor
for the measuring process, e.g. very slow placing of a load on the weighing pan
(load adding, e.g. pouring, filling). In such case, it is recommended to disable
the function.