19.2.1. Working Mode Accessibility
To enable/disable given working mode, press key.
Procedure:
Enter <P3.Func> menu and select given working mode.
Go to <Acc> function.
Press key, parameter values are displayed successively one by one:
YES – working mode enabled, no – working mode disabled.
Press key to confirm, next go to the home screen.
19.2.2. Save Mode
Parameter allowing you to set mode of sending data from the weighing device
to a peripheral device.
Procedure:
Enter <P3.Func> menu and select given working mode.
Go to <Snn> function.
Press key, parameter values are displayed successively one by one:
Manual printout of stable weighing result. Upon pressing key at the moment
when the result is unstable (no pictogram displayed), the program first waits for
the stability condition to be met, only then printout is carried out.
Manual printout of each weighing result. In case of unstable indication, <?> sign
is displayed in front of the 'mass frame'. Function available for non-verified
scales exclusively.
Automatic printout of the first stable weighing result above <Lo> threshold (to set
<Lo> threshold go to <Lo> parameter).
Automatic printout with time interval set in [min] (to set the interval go to <Int>
parameter).
Press key to confirm, next go to the home screen.
19.2.3. Automatic Printout Time Interval
Parameter enabling you to set frequency of automatic printout. Printout interval
is set in minutes with 1 [min] accuracy within 1 [min] - 1440 [min] range.