pH/Cond 3320 pH value
ba77054e03 03/2018 27
Continuing with
three- to five-point
calibration
13. When measuring without temperature sensor:
Temper the buffer solution, or measure the current temperature.
Enter the temperature value in the menu.
14. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
15. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
16. If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
pH
0.0
24.8 °C
AutoCal TEC
7.000
15.02.2014
14:15
AR
Buffer
mV
17. Thoroughly rinse the electrode with deionized water.
18. Immerse the electrode in the next buffer solution.
19. When measuring without temperature sensor:
Temper the buffer solution, or measure the current temperature.
Enter the temperature value in the menu.
20. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.