8.2.4 Calibration data
Displays the calibra-
tion data
The calibration record of the last calibration is to be found under the menu item,
<ENTER> / Calibration / Calibration record. To open it, press the <CAL
_ > key
in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the <F2>[USB output] key.
Displaying the cali-
bration data memory
The calibration records of the last calibration procedures are available in the
<ENTER>/Calibration / Calibration data storage and <ENTER>/Storage &
config / Data storage / Calibration data storage menu.
4. Immerse the conductivity measuring cell in the control standard solu-
tion, 0.01 mol/l KCI.
5. Start the measurement with <ENTER>.
The [AR] status indicator is displayed. A progress bar is displayed and
the display of the measured parameter flashes.
6. Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration record is displayed and output to the interface.
7. Using <F1>/[Continue] or <ENTER>, switch to the measured value
display.
The calibration record is automatically transmitted to the interface
after calibrating.
ϰ
0.479 1
/cm
25.1 °C
15.02.2014
08:00