pH/Cond 3320 Settings
ba77054e03 03/2018 53
* Brand names or trade names are trademarks of their respective owners protected
by law.
9.1.3 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
17 Beckman * 4.005
7.005
10.013
25 °C
18 Hamilton Duracal * 4.005
7.002
10.013
25 °C
19 Precisa * 3.996
7.003
8.999
25 °C
20 Reagecon TEC * 2.000
4.010
7.000
10.000
25 °C
21 Reagecon 20 * 2.000
4.000
7.000
10.000
13.000
20°C
22 Reagecon 25 * 2.000
4.000
7.000
10.000
13.000
25 °C
23 Chemsolute *2.000
4.000
7.000
10.000
20°C
24 USABlueBook
*4.000
7.000
10.000
25 °C
25 YSI * 4.000
7.000
10.000
25 °C
No. Buffer set * pH values at
The buffers are selected in the menu, pH / <ENTER> / Calibration
/ Buffer (see 9.1.1 S
ETTINGS FOR PH MEASUREMENTS, PAGE 50).