The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the <ENTER> (or <F1>/[Menu]) key.
Cond The following settings for conductivity measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the <ENTER> (or <F1>/[Menu]) key.
9.3.2 Resetting the system settings
The following system settings can be reset to the default status:
Slope -59.2 mV
Man. temperature 25 °C
One point calibration No
Setting Default settings
Setting Default settings
Kal.-Intervall 150 d
Measured parameter
ϰ
Cell constant (C)
(calibrated)
0.475 cm
-1
or 0.840 cm
-1
(nominal cell constant of the conduc-
tivity measuring cell last calibrated)
Cell constant (C)
(set)
0.470 1/cm
Temperature compensation nLF
Reference temperature 25 °C
Temperature coefficient (TC) of the
linear temperature compensation
2.000 %/K
TDS multiplier 1.00
Setting Default settings
Language English
Beep On
Baud rate 4800 Baud
Output format ASCII
Decimal separator Dot (xx.x)