pH/Cond 3320 What to do if...
ba77054e03 03/2018 77
13 What to do if...
13.1 pH/ORP
Error message
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Electrode:
Air bubble in front of the junction Remove air bubble
Air in the junction Extract air or moisten junction
Cable broken Replace the electrode
Gel electrolyte dried out Replace the electrode
The measured value is outside
the measuring range of the meter
Use a suitable electrode
Cause Remedy
Electrode:
The values determined for zero
point and slope of the electrode
are outside the allowed limits.
Recalibrate
Junction contaminated Clean the junction
Electrode broken Replace the electrode
Buffer solutions:
Incorrect buffer solutions Change calibration procedure
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions