50 ba77054e03 03/2018
9 Settings
9.1 Measurement settings
9.1.1 Settings for pH measurements
The settings for pH measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the <ENTER> key. After completing the settings, switch to the measured
value display with <M>.
In the following table, only those settings are listed that concern the pH mea-
surement.
Default settings are printed in bold.
Menu item Possible set-
ting
Explanation
Calibration /
Calibration record
- Displays the calibration record of
the last calibration.
Calibration /
Calibration data
storage / Display
- Displays the calibration record.
Further options:
Scroll through the calibration
records with <>< >.
Output the displayed calibration
record to the interface with
<F2>/[USB output].
Output all calibration records to
the interface with <F2
_ >/[USB
output].
Quit the display with <F1>/
[Back] or <ENTER>.
Switch directly to the measured
value display with <M>.
Calibration /
Calibration data
storage / Output to
USB
- Outputs the calibration records to
the interface.
Calibration / Buffer TEC
NIST/DIN
ConCal
...
Buffer sets to be used for pH cali-
bration (see section 5.2 C
ALIBRA-
TION, page 25).
Calibration / One
point calibration
Yes
No
Quick calibration with 1 buffer