pH/Cond 3320 Technical data
ba77054e03 03/2018 85
14.2 General data
Accuracy (± 1 digit) Variable Accuracy Temperature of the test
sample
ϰ and ρ / temperature compensation
None (Off) ± 0.5 %
Nonlinear (nLF) ± 0.5 % 0 °C ... +35 °C
according to EN 27 888
± 0.5 % +35 °C ... +50 °C
enhanced nLF function
Linear (lin) ± 0.5 % +10 °C ... +75 °C
SAL / range
0.0 ... 42.0 ± 0.1 +5 °C ... +25 °C
± 0.2 +25 °C ... +30 °C
TDS [mg/l]
± 0.5 %
T [°C] / temperature sensor
NTC 30 ± 0.1
PT 1000 ± 0.1
The accuracy values specified here apply exclusively to the meter.
The accuracy of the measuring cell has also to be taken into
account.
Further data are given in the documentation of your sensor.
Dimensions ca. 180 x 80 x 55 mm
Weight Approx. 0.4 kg
Mechanical structure Type of protection IP 67
Electrical safety Protective class III
Test certificates CE
Ambient
conditions
Storage -25 °C ... +65 °C
Operation -10 °C ... +55 °C
Admissible relative
humidity
Yearly mean: < 75 %
30 days/year: 95 %
Other days: 85 %