Settings pH/Cond 3320
60 ba77054e03 03/2018
Application tips Set the temperature compensation suitable for your test sample:
9.1.9 Setting the TDS factor
The factor to calculate the total dissolved solids is set to 1.00 in the delivery
condition.
You can adjust this factor to meet your requirements in the range 0.40 ... 1.00.
The factor is set in the menu for the parameter TDS (Measurement / TDS fac-
tor).
The reference temperature and temperature compensation are set
in the Measurement menu for the parameter, conductivity (see sec-
tion 9.1.6 S
ETTINGS FOR CONDUCTIVITY MEASURING CELLS, page 57).
Test sample Temperature compensation Display
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water nLF
according to EN 27 888
nLF
Other aqueous solutions Lin
Set linear temperature coeffi-
cient 0.000 ... 10.000 %/K
Lin
Salinity (seawater) Automatic nLF according to
IOT (International Oceano-
graphic Tables)
Sal, nLF