Transmitting data (USB interface) pH/Cond 3320
72 ba77054e03 03/2018
11 Transmitting data (USB interface)
11.1 Options for data transmission
Via the USB interface you can transmit data to a PC. The following table shows
which data are transmitted to the interface in which way:
Data Control Operation / description
Current datas-
ets of the sen-
sors indicated
on the display
Manual With <F2>/[USB output].
Simultaneously with every manual
data storage process (see section
10.1 M
ANUAL STORAGE, page 66).
Automatic, at
intervals
With <F2
_ >/[USB output].
Then you can set the transmission
interval.
Simultaneously with every auto-
matic data storing process (see
section 10.2 A
UTOMATIC DATA
STORAGE AT INTERVALS, page 67).
Stored mea-
sured values
Manual Displayed dataset with <F2>[USB
output] after calling up from the
memory.
All datasets with the Output to USB
function.
(see section 10.3.1 E
DITING THE
MEASUREMENT DATA MEMORY, page
69).
Calibration
records
Manual Calibration record with <F2>/[USB
output]
(see section 5.2.6 C
ALIBRATION
DATA, page 32).
Automatic At the end of a calibration proce-
dure.
The following rule applies: With the exception of the menus, shortly
pressing the <F2>[USB output] key generally outputs the display
contents to the interface (displayed measured values, measure-
ment datasets, calibration records).