pH/Cond 3320 Conductivity
ba77054e03 03/2018 49
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration record
(example)
Menu item Setting/func-
tion
Explanation
Calibration /
Calibration data stor-
age / Display
or
Data storage /
Calibration data stor-
age / Display
- Displays the calibration record.
Further options:
Scroll through the calibration
records with <>< >.
Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].
Quit the display with <F1>/
[Back] or <ENTER>.
Switch directly to the mea-
sured value display with <M>.
Calibration /
Calibration data stor-
age / Output to USB
or
Data storage /
Calibration data stor-
age / Output to USB
- Outputs the calibration records to
the interface.
Display Calibration
record
Cell constant [cm
-1
]
+++ within the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
Error Error outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
Error elimination (see section 13
W
HAT TO DO IF..., page 77)
pH/Cond 3320
Ser. no. 11292113
CALIBRATION Cond
15.02.2014 16:13:33
Cell constant 0.479 1/cm 25.0 °C
Sensor +++