Ion concentration pH/Cond 3320
44 ba77054e03 03/2018
Displaying the cali-
bration data memory
The calibration records of the last calibrations are available in the menu,
<ENTER> / Calibration / Calibration data storage.
Calibration evalua-
tion
After calibrating, the meter automatically evaluates the calibration.
Calibration record
(example)
Menu item Setting/func-
tion
Explanation
Calibration /
Calibration data stor-
age / Display
- Displays the calibration record.
Further options:
Scroll through the calibration
records with <>< >.
Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].
Quit the display with <F1>/
[Back] or <ENTER>.
Switch directly to the mea-
sured value display with <M>.
Calibration /
Calibration data stor-
age / Output to USB
- Outputs the calibration records to
the interface.
Display Calibration
record
Magnitude of the slope [mV]
+++ 50.0 ... 70.0 or 25.0 ... 35.0
Error Error < 50 or > 70
or
< 25 or > 35
Error elimination (see section 13 W
HAT TO
DO IF..., page 77)
pH/Cond 3320
Ser. no. 12345678
CALIBRATIONISE
18.01.2013 08:09:10
Standard 1 0.010 mg/l
Standard 2 0.020 mg/l
Voltage 1 38.5 mV
Voltage 2 58.0 mV
Temperature 1 24.0 øC
Temperature 2 24.0 øC
Ion type Ag
Slope 64.7 mV
Sensor
+++
_____________________________________