Settings pH/Cond 3320
58 ba77054e03 03/2018
Measurement /
Measuring cell /
Type
Measuring cell used
Cal Measuring cells whose cell constant
is determined by calibration in the
KCL control standard solution.
Calibration ranges:
0.450 ... 0.500 cm
-1
and
0.800 ... 0.880 cm
-1
The currently valid cell constant is
displayed in the status line.
LR 325/01
Measuring cell LR 325/01,
nominal cell constant
0.100 cm
-1
.
The cell constant can be adjusted in
the range 0.090 ... 0.110 cm
-1
.
LR 325/001 Measuring cell LR 325/001,
nominal cell constant
0.010 cm
-1
.
The cell constant is permanently set.
man Any measuring cells with freely
adjustable cell constants.
Measurement /
Measuring cell /
Cell const. man
0,250 ...
0,475
25,000 cm
-1
Display and setting option of the cell
constant of any measuring cells
(man).
Measurement /
Measuring cell /
Cell const. LR325/
01
0.090 ...
0.100 ...
0.110 cm
-1
Display and setting options for the cell
constant of the measuring cell
LR 325/01.
Measurement /
Temp. comp. (TC)
/ Method
nLF
Lin
Off
Procedure for temperature compen-
sation (see section 9.1.8 T
EMPERA-
TURE COMPENSATION, page 59).
This setting is only available for the
measured parameters,
ϰ and ρ.
Measurement /
Temp. comp. (TC)
/ Linear coeff.
0.000 ...
2.000 ...
10.000 %/K
Coefficient of the linear temperature
compensation.
This setting is only available when the
linear temperature compensation is
set.
Menu item Possible
setting
Explanation