Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
21
6. Refer to the remaining Total, Direct, or R/T Calibration procedures that follow.
Total Calibration
1. Place the sample transmission holder base inside the transmission compartment.
2. Attach the appropriate clamp and stop to the base with the thumb screws.
3. Center the black plastic blocking panel in the transmission sample holder and place between
the stop and clamp toward the sphere.
4. Close the cover and continue with calibration.
Black plastic
blocking panel
Carriage base
Total transmission
calibration shown