Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
23
5. Once the calibration process is complete, the calibration LED becomes lit. Any change to the
spectrophotometer configuration may result in the calibration LED turning from green
(calibrated) to red (not calibrated). Remember that each configuration needs to be calibrated.
R/T Calibration
1. Place the sample transmission holder base inside the transmission compartment.
2. Attach the appropriate clamp and stop to the carriage base with the thumb screws.
3. Center the black plastic blocking panel in the transmission sample holder and position toward
the sphere.
4. Close the cover and continue with calibration.
5. Once the calibration process is complete, the calibration LED becomes lit. Any change to the
spectrophotometer configuration may result in the calibration LED turning from green
(calibrated) to red (not calibrated). Remember that each configuration needs to be calibrated.
Note: The green LED will never appear for any Calibration Mode configured to include a UV Cal
option. UV cannot be calibrated in this mode.
blocking panel