Number Item
Setting
value
Description of options
(The underlined option is the factory setting.)
#09
Gravity compensation for the
region of use
0:
Parameter for instruments for other than transactions or
certifications
#22 Weighing speed 0: Standard
1:
Speed-oriented
* If you specify this option for an instrument for transactions or
certifications, it becomes invalid (0: standard).
2: Accuracy-oriented
#25 LED brightness 0: 25% of full brightness
1: 50% of full brightness
2: 75% of full brightness
3: 100% of full brightness
#P3
Unit of proper weight value for
checkweighing function
0:
Scale interval
1: Ratio with respect to the target value (%)
2: Ratio with respect to the weighing capacity (%)
#P4
Proper weight value for
checkweighing function
0 to 255: Range in units specified in #P3 (Factory setting: 10)
5-4. Description of each user parameter
[#01] Checkweighing function
Select a type of the checkweighing function built in the Scalapro
you wish to use.
[#03] LED lighting timing
Select the timing of turning on the Evaluation LED light when the checkweighing function is used.
[#05] Automatic power OFF time
You can automatically turn off the power of the instrument when it is not used for a specified period
(minutes).
[#07] Flashing of indicated value per evaluation (parameter for checkweighing function)
When the checkweighing function is enabled, you can make a setting so that the indicated value flashes. For
example, if you set it to flashing when the weighed items have the proper weight, the evaluation result will be
clearer.
[#09] Gravity compensation for the region of use
Make this setting when you use the instrument for other than transactions or certifications in another
gravitational region. Refer to page 21 for details.
[#22] Weighing speed
You can increase the weighing speed; or you can decrease the weighing speed to ensure accurate
measurements when you weigh items whose measured value takes a long time to be stable.
[Note] Be aware that you cannot increase the weighing speed when you use an instrument for transactions
or certifications.
[#25] LED brightness
You can change the brightness of the LED light used for the checkweighing function.
[#P3] Unit of proper weight value for checkweighing function
From the following options, you can select a unit of the proper weight value for the checkweighing function:
1: Scale interval,
2: Ratio with respect to the target value (%),
3: Ratio with respect to the weighing capacity (%).
[#P4] Proper weight value for checkweighing function
Set the proper weight range in units specified in #P3.
If you set #P3 to “3” and #P4 to “10,” the proper weight range will be 10% of the weighing capacity.