7-4. How to conduct weight calibration
For the Scalapro
™
for a purpose other than transactions or certifications, the user can conduct weight
calibration (calibration before use). It is recommended to periodically conduct weight calibration because
otherwise measurement results may become inaccurate. Get Class 2 or better standard weights and conduct
weight calibration according to the following procedure:
[Example] Procedure for weight calibration when you use an instrument with a weighing capacity of 60 kg:
* Before conducting weight calibration, ensure to put a pre-load equivalent to the weighing capacity on the
instrument.
Operation Display
Press
and at the same time to enter the user
parameter setting mode.
Press
and at the same time while the parameter #01
is displayed.
The weight calibration screen appears. Press to start a
calibration.
Put 30kg weight(s), which is equivalent to 3000 scale intervals, and
then press
.
* Put 15kg weight(s) when the weighing capacity is 32kg, or put 60kg
weight(s) when the weighing capacity is 120kg.
Put 60kg weight(s), which is equivalent to 6000 scale intervals, and
then press
.
* Put 30kg weight(s) when the weighing capacity is 32kg, or put 120kg
weight(s) when the weighing capacity is 120kg.
The weight calibration is now complete. Upon completion, the panel
returns to mass indications. Put the weight(s) again to verify that the
value indicated on the panel matches the actual mass of the weight(s).
* To cancel the calibration during the process, hold down
.