21
4.
OPERATION PROCEDURE
Time & Temperature Settings
Setting sterilization mode temp/time
◆ Set the sterilization temp/time.
◆ Simply press “MODE” to cancel settings at any point during setup.
◆ Screen automatically reverts to standby (no settings saved) if keys go unpressed for more than a minute
during set up.
Standby
Turn main switch ON.
Main display flashes sterilization temperature. Sub display flashes
sterilization time.
Sterilization Temperature
Press “MODE”.
Main display flashes . Sub display flashes sterilization
temperature.
Press “ENTER”. Sterilization temperature can now be set.
Enter desired sterilization temperature using ▽△.
▽…decreases temperature value.
△…increases temperature value.
Press “ENTER”.
Sterilization temperature setting is finalized. Displays show sterilization
time setting.
Sterilization Time
Press “ENTER” again.
Sterilization time can now be set.
Enter desired sterilization time using ▽△.
▽…decreases time value.
△…increases time value.
Press “ENTER”.
Sterilization time setting is finalized. Displays show drying temperature
setting.
Press “MODE” to exit setup menu at this point, if drying mode will not
be used. Otherwise, press “ENTER” and proceed to the next step.
Setting drying mode temp/time.
Drying Temperature
Enter desired drying temperature using ▽△.
▽…decreases temperature value (min.150°C).
△…increases temperature value (max.180°C).
Press “ENTER”.
Drying temperature setting is finalized. Displays show drying time
setting.