8. TROUBLESHOOTING
Troubleshooting Guide
Troubleshooting table
Symptom Check Points
Display does not come up when
main power switch is turned on.
Is the power cable connected?
Is a power outage in progress?
Is power source voltage adequate?
Air does not purge.
Safety valve is activated.
Is the exhaust hose kinked or clogged?
Is the line clogged with debris?
Water does not drain.
Water does not drain in drying
process.
Is drain filter clogged?
Sterilization temperature does rise.
Pressure does not build.
Is the lid closed securely?
Is seal or flange damaged?
Pressure increases when solenoid
valve is open.
Is the chamber exhaust port clogged?
Temperature fluctuates during
operation.
Are there considerable changes in external temperature?
Steam discharges rapidly.
Is there water in the drain bottle?
Is exhaust hose positioned incorrectly or damaged?
Water leaks.
Is drain valve closed all the way?
Is the drain bottle overfilled?
Operation does not start from
stand-by.
Is drain bottle in position?
Loud noise during air purge. Is the silencer installed correctly?
For further technical assistance, contact Yamato Scientific at (800) 292-6286