EasyManua.ls Logo

Yamato SM511 - Sterilize Operation Process

Yamato SM511
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
4. OPERATION PROCEDURE
Sterilize
STERILIZE operation process
Confirm the following:
Is the drain valve closed? P.8
Is the drain filter installed? P.7
Is the bottom plate in place? P.7
Is the drain bottle in place? P.9
Does chamber have proper amount of water? P.10
Is the lid closed completely? Pg.11
Setup
Turn on main power switch.
Press “MENU” key and select “STERILIZE” (See P.22).
Set the sterilization temperature and time as desired
(See P.21).
Main display flashes temperature setting.
Sub display flashes time setting.
Sterilize menu lamp flashes.
Operation
Start / Air Purge
Press “ENTER”.
Operation begins. Sterilization heater turns on.
The solenoid exhaust valve opens. Air is purged from chamber.
(Approximately 20-25 minutes)
Main display shows water temp and heater lamp status.
Sub display remains blank.
Sterilize process lamp (right) flashes.
Sterilize menu lamp (left) lights.
Pressurization
When air purge stage is completed, the solenoid exhaust valve
closes, and pressurization stage begins.
(Approximately 20-25 minutes)
Main display shows chamber temperature.
Sterilize process lamp lamp flashes.
Sterilize menu lamp lights.
Sterilization
When the chamber temperature reaches temperature setting,
sterilization process begins and the timer begins counting.
At this point, sub display shows remaining sterilization time.
Main display shows chamber temperature.
Sub display shows remaining time.
Sterilize process lamp is lights.
Sterilize menu lamp is lights.

Related product manuals