3-2
IM CA500-01EN
3.1 DC Voltage Measurement
Description
Measurement Range
You can select from the following three measurement ranges.
Range Measurement Range
100 mV ±110.000 mV
5 V ±6.0000 V
50 V ±55.000 V
0% Value and 100% Value
Assign an output value (specified according to the specifications of the device to be calibrated) to
the 0% or 100% measurement value of this instrument, which is mapped to the 0% or 100% source
value of this instrument.
The instrument displays errors, pass/fail judgment results (CA550), measurement percentages
relative to the assigned value.