Page #122 background imageLoading...
Page #122 background image
7-19
IM CA500-01EN
1
2
3
4
5
6
7
Index
USB Function
7.5 Error Codes
Indication Description
Err 00 No error (not displayed on the screen)
Err 11 Received an unused command
Err 12 Incorrect command parameter designation
Err 13 Received a command that cannot be executed in the
instrument’s current condition
Err 16 Detected an error during adjustment
Err 20 24 VLOOP measurement power supply error
Err 23 Detected an overvoltage or overcurrent in the source
output
Err 60 Inappropriate setting information saved in EEPROM
Err 61 Inappropriate measurement adjustment information saved
in EEPROM
Err 62 Inappropriate source adjustment information saved in
EEPROM