Parameter setting
Quick Reference Instruction Manual for Spare
Transmitter replacement
IM01U10A01-00EN-R, 1
st
edition, 2019-12-09
45 / 76
7.3.4 Set up sensor relevant and sensor specific parameters
For Spare
transmitter:
Set the Spare transmitter sensor parameters to the values from microSD card of the de-
fective transmitter by display.
For Spare
sensor:
Set the sensor parameters in the existing transmitter to the values from microSD card of
the Spare sensor by display.
1. Set user level “Specialist”
2. Go to menu item “[Diag/Service] ► [Param bkup/restore] ► [Restore]”
3. Choose sub menu item “[Bkup name]”. Put in the backup name of the defective trans-
mitter user configuration file; for example “UsrPlant”
4. Go to menu item “[Diag/Service] ► [Param bkup/restore] ► [Restore]”
5. Execute “[Download sensor constant from microSD]”
6. Check all sensor relevant settings of the exchange transmitter
7.3.5 Set up customer specific parameters
The default setting for customer specific parameters is as follows:
Tab.6: Assignment: default setting for customer specific parameters
Parameter name: Default Setting: Parameter
adjustment:
Value Unit
[Mass flow lowcut]
0.0105 kg/h
Recommended
if meter size is >06
[Net mass flow 1 lowcut]
1)
[Net mass flow 2 lowcut]
1)
[Mass flow URV]
21.000 kg/h
Recommended
if meter size is >06
[Net mass flow 1URV]
1)
[Net mass flow 2URV]
1)
[Volume flow URV] 21.000 L/h
[Liquid gas select] Liquid None
Needed
if sensor is a gas device
[Max permissible pressure] 10.000 Bar
Automated
with restoration flow
1)
Relevant only if options AC␣ are available for the Spare transmitter