What to do if I get Alarm 7 on my YOKOGAWA ZR402G Measuring Instruments?
- JJohn RobinsonAug 14, 2025
If you're seeing Alarm 7, it could be due to a discrepancy between the oxygen concentration of the span gas and the set value, or the span gas flow might be outside the specified range (600 ± 60 ml/min). In such cases, try performing the calibration again. Additionally, a damaged cell assembly with abnormal cell voltage can trigger this alarm, requiring a cell replacement.