<9. Calibration>
9-10
IM 11M12A01-04E 11th Edition : Jul. 19, 2017-00
9.3.3 Automatic Calibration
No execution operations are required for automatic calibration. Automatic calibration starts in
accordance with a preset start day and time. Calibration is then executed at preset intervals.
NOTE
Before conducting a semi-automatic or automatic calibration, run the automatic calibration unit
beforehand to obtain a calibration ow of 600 ± 60 ml/min.