OPERATION
ZEISS TFT display touchscreen on the Axio Observer 7, 7 materials Axio Observer
120 431004-7244-001 12/2016
TL / RL illumination switch
The Off and On switches open or close the shutter for reflected light (RL) and transmitted light (TL) or
switch the illuminators on and off.
Definite Focus switch
The Off and On switches switch Definite Focus on and off.
If the system does not have Definite Focus, the Make it visible! Button will appear in place of
the Off and On switch.
Make it visible! button
If the microscope is so poorly adjusted that no image of the specimen is visible, this button can be used
to reset the microscope to a standard state in which the specimen is visible.
The following settings are adapted in this basic state:
Bio / Med stand:
transmitted light illuminator is set to medium brightness (3 V)
aperture diaphragm is opened
TL shutter open, RL shutter closed
Motorized condenser is switched to brightfield
reflector turret is rotated to the nearest brightfield position
Light path switching is set to 100 % vis
MAT (materials) stand:
Reflected light illuminator (HAL 100 or LED) set to medium brightness
TL shutter closed
RL shutter opened
Reflector turret in brightfield reflected light position (alternatively contrast module)
reflected light luminous-field diaphragm opened
reflected light aperture diaphragm opened
Optovar turret in position 1: Optovar 1x
Sideport in position 1 (100% vis)
Baseport in position 2 (100% vis)
The nosepiece position (and, if entered, the objective) will be shown on the left of the status bar, the
halogen illuminator voltage on the right.