Operating Manual
Axio Examiner
Research Microscope

Table of Contents

Questions and Answers

  • P
    Paul MyersAug 15, 2025
    What should I do if my Zeiss Microscope lamp won't turn on?
    • C
      Chase Cannon MDAug 15, 2025
      If the 12 V 100 W halogen lamp in your Zeiss Microscope doesn't light up even when the on/off switch is activated, first ensure the power plug is properly connected to the mains outlet, verifying both equipment and line voltage. If it still doesn't work, check if the 12 V 100 W halogen lamp is installed correctly. If not, insert it. If the lamp is installed, it may be defective and require replacement. Confirm that you are using the mandatory 12 V 100 W halogen lamp. Lastly, check and replace the fuses if necessary.
  • P
    Paul CarpenterAug 18, 2025
    Why is my Zeiss Axio Examiner Microscope image dark or unevenly lit?
    • L
      lisasmithAug 19, 2025
      Uneven brightness or shading in the Zeiss Microscope's visual field can occur for several reasons. The vis/fot push rod at the phototube might not be in the correct functional position; ensure it's switched to the final position. Check that the objective is fully lowered to the working position. Also, the condenser, aperture stop, or luminous-field diaphragm might not be set correctly; adjust them as needed. Lastly, confirm that the filter is correctly placed in the filter holder.
  • S
    Shane HansonAug 21, 2025
    Why is my Zeiss Microscope image blurry and has low contrast?
    • L
      lewisrobertAug 21, 2025
      Low resolution and poor image contrast in your Zeiss Microscope could stem from several factors. Ensure the aperture stop opening is correctly adjusted. Focus the condenser. For transmitted light objectives, use standardized cover slips with a thickness of 0.17 mm. If using immersion objectives, apply Immersol 518 N® or 518 F® from Carl Zeiss, removing any air bubbles. Clean any immersion oil from dry objective lenses. Adjust the correction setting ring to match the cover slip thickness. Finally, clean any dirt or dust from the optical surfaces of the objectives, eyepieces, condensers, or filters.
  • R
    Richard MendezAug 25, 2025
    What to do if my Zeiss Axio Examiner Microscope image is blurry on one side?
    • F
      fdillonAug 25, 2025
      If you're experiencing asymmetric image definition with your Zeiss Microscope, where one side is in focus and the other is not, first ensure the condenser is correctly set. Next, make sure the objective is fully lowered to its working position. Finally, check that the specimen is correctly inserted and clamped in the specimen holder on the mechanical stage.
  • S
    Savannah BakerAug 27, 2025
    Why is my Zeiss Axio Examiner tiring my eyes?
    • R
      ryan22Aug 27, 2025
      Eye strain during microscopy with your Zeiss Microscope can be due to several reasons. Check that the eyepiece distance (interpupillary distance) is correctly set on the binocular tube. Ensure the focusable eyepieces are adjusted to your individual eye deficiencies. Adjust the lamp voltage or insert a conversion filter to achieve acceptable image brightness. If these steps don't resolve the issue, the binocular tube may be maladjusted optically or mechanically.
  • C
    Carolyn WhiteAug 30, 2025
    How to fix major focus differences when changing objectives on my Zeiss Axio Examiner Microscope?
    • N
      Nicholas RodriguezAug 30, 2025
      If you notice significant focus differences when switching objectives on your Zeiss Microscope, first adjust the focusable eyepieces to correct for any eye deficiencies. Next, ensure each objective is fully screwed in up to the stop face. Lastly, verify that the tube lens is either correctly installed or that a redundant tube lens is removed.
  • A
    andrea78Sep 2, 2025
    How do I get rid of spots in my Zeiss Axio Examiner's visual field?
    • T
      teresaboyleSep 2, 2025
      To eliminate dirt or dust in the visual field of your Zeiss Microscope, focus the condenser and correctly switch the 0.9 front lens on or off. Also, adjust the aperture stop opening if it's too small. Finally, clean the optical surfaces of the objectives, eyepieces, condensers, filters, or specimens to remove any dust or debris.
  • A
    Adam JonesSep 5, 2025
    Why can't I merge the images from both eyepieces on my Zeiss Microscope?
    • R
      Regina RhodesSep 5, 2025
      If you're unable to combine the left and right visual fields into a single image with your Zeiss Microscope, first adjust the eyepiece distance (interpupillary distance) correctly on the binocular tube. Also, ensure that the focusable eyepieces are properly adjusted to correct for any differences in your vision.
  • M
    melissajonesSep 8, 2025
    Why is the halogen lamp flickering in my Zeiss Axio Examiner?
    • C
      christopherowenSep 8, 2025
      If the 12 V 100 W halogen lamp in your Zeiss Microscope is flickering or its luminosity is unstable, it could be nearing the end of its service life and needs replacement. Ensure the AC power cord is correctly installed and not damaged. Also, verify that the pins of the 12 V 100 W halogen lamp are properly inserted into the socket.
  • D
    Diana RogersSep 12, 2025
    What to do if the motorized part of my Zeiss Microscope is unresponsive?
    • E
      ecarterSep 12, 2025
      If the motor-operated lower part of your Zeiss Microscope isn't responding, first check that the microscope is connected to its control box. Then, ensure the control box is switched on and properly connected. Verify that the docking station is also connected. Finally, confirm that the motors are switched on by checking if the blue indicator light is lit.

Summary

1 INTRODUCTION

1.1 Equipment safety notices

Notices and warnings for safe operation and potential hazards of the equipment.

1.2 Warning and information signs at the equipment

Explanation of warning symbols and safety signs used on the equipment.

3 START-UP

3.1 Installing standard components

Step-by-step guide for unpacking and setting up the basic microscope components.

3.2 Establishing a mains connection, and switching the power supply units on or off

Instructions for connecting power and operating the microscope and its illuminators.

4 OPERATION

4.1 Illumination and contrast procedures

Methods for optimizing illumination and contrast for various microscopy techniques.

5 CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE

5.1 Equipment care

Guidelines for cleaning, storage, and environmental considerations for the microscope.

5.3 Trouble shooting

Common problems, their causes, and recommended solutions for troubleshooting issues.

Zeiss Axio Examiner Specifications

General IconGeneral
TypeUpright Microscope
IlluminationLED or halogen
StageMotorized or manual stage options
FocusingCoarse and fine focus knobs
Focus DriveMotorized or manual
Camera IntegrationYes
SoftwareZEN (Zeiss Efficient Navigation)
Contrast MethodsBrightfield, Darkfield, Phase Contrast, DIC, Fluorescence
Optical SystemInfinity Optical System