EasyManua.ls Logo

Zeiss Axio Examiner - Setting the Lock-In Resistance of the 5 X Reflector Turret; Setting the Actuating Rod

Zeiss Axio Examiner
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
START-UP
Axio Examiner Microscope basic setting Carl Zeiss
M60-2-0003 e 05/2012 57
3.4.4 Setting the lock-in resistance of the
5x reflector turret
Only for Axio Examiner.D1 and .Z1 micro-
scope top parts.
The lock-in resistance of the 5x reflector turret is
freely adjustable. Proceed as follows:
Remove the binocular tube (see section 3.1.3).
Pull the closing plug (Figure 3-46/1) out of the
upper microscope part.
Turn the adjusting screw (Figure 3-46/2) using a
ball head screwdriver until the desired lock-in
resistance is set.
Insert the closing plug again (Figure 3-46/1).
Put on the binocular tube (see section 3.1.3).
3.4.5 Setting the actuating rod
Only for phototubes 425521-9020-000 and
425521-9000-000.
You can adjust the motion of the actuating rod
used for toggling between eyepieces and camera
output. Proceed as follows:
Remove the phototube from the upper
microscope part (see section 3.1.3).
Lay down the phototube safely with the lower
side turned upwards.
Set the motion step using the screw (width
A/F 1.5) (Figure 3-47/1) located on the lower
side.
Put the phototube on the upper microscope
part (see section 3.1.3).
Figure 3-46 Setting the lock-in resistance
Figure 3-47 Setting the actuating rod

Table of Contents

Related product manuals