EasyManua.ls Logo

Zeiss Axio Examiner - Figure 3-19 Adjusting the HAL 100 Halogen Lamp

Zeiss Axio Examiner
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
START-UP
Carl Zeiss Installing standard components Axio Examiner
38 M60-2-0003 e 05/2012
Adjusting the HAL 100 halogen 3.1.13.2
lamp
(1) Coarse adjustment
Loosen the clamping screw (Figure 3-18/2), and
remove the ready-to-operate halogen
illuminator from the microscope stand.
Switch on the 12 V DC 100 W external power
supply unit (see section 3.2.2).
Direct the light beam against a projection
surface (wall) which is at least 3 m away.
Do not look into the illuminator's light
exit opening.
Set the adjusting screw (Figure 3-19/3) using a
ball head screwdriver (width A/F 3) to reach the
best focus possible for both images of the lamp
filaments on the projection surface.
Then, set the adjusting screws (Figure 3-19/4
and 5) to a value at which the lamp filaments of
one image exactly fill out the gaps in the
reflector image (Figure 3-19/1).
(2) Fine adjustment
Attach the microscope lamp and locate it in position using the clamping screw at the microscope
stand again.
If required, switch out the transmitted-light diffusing screen, and remove the filter drawer from the
corresponding holder compartment.
Focus the 40x lens on the specimen, and find a free object position.
Remove the eyepiece, and center the lamp filament and the lamp filament's mirror image in the pupil
image using the adjusting screws (Figure 3-19/4 and 5).
Set the adjusting screws (Figure 3-19/3) to optimize the uniform illumination of the pupil image.
If required, switch on the diffusing screen, and insert the filter drawer again.
Figure 3-19 Adjusting the HAL 100 halogen
lamp

Table of Contents

Related product manuals