EasyManua.ls Logo

Zeiss Axio Examiner - Installing the Aqua Stop for the Condenser; Attaching a Condenser Carrier with a Motor-Operated Focus

Zeiss Axio Examiner
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
START-UP
Axio Examiner Installing optional components Carl Zeiss
M60-2-0003 e 05/2012 55
3.3.10 Installing the Aqua Stop for the
condenser
Remove the stage carrier (see section 3.1.12).
Connect the discharge hose (Figure 3-42/2) to
the Aqua Stop (Figure 3-42/1).
Carefully put the Aqua Stop (Figure 3-42/1) on
the condenser (Figure 3-42/3). The rim must fit
in the slot at the condenser.
Run the discharge hose (Figure 3-42/2) to a
corresponding vessel (dish, etc.).
Put on the stage carrier (see section 3.1.12).
3.3.11 Attaching a condenser carrier with
a motor-operated focus
To install a motor-operated condenser carrier,
follow the same procedure as for a manual
condenser carrier (see section 3.1.9).
After assembly, connect the condenser carrier
control cable (Figure 3-43/1) to the lower
microscope part.
Figure 3-42 Installing the Aqua Stop
Figure 3-43 Connecting a mot. condenser carrier

Table of Contents

Related product manuals