3.1 General
Prior to using the device on a patient, the user should become familiar with the instructions for
use and individual treatment methods to be used as well as the indications / contraindications,
warnings and application information. Additional sources of information about the ultrasound
therapy should be followed.
These instructions for use must always be stored with the device and kept accessible at all
times for anyone authorised to operate this device.
To avoid the risk of electric shock, the plug must be disconnected from the power supply before
performing any cleaning or maintenance activities.
Before use, ensure that the device is powered via a properly grounded mains socket (electrical
installation according to DIN VDE 0100 Part 710). The device must only be operated with the
supplied power cable. The power cable must be protected against mechanical stress.
Magnetic and electrical fields can affect the function of the device. For this reason, do not
operate the device in the vicinity of devices which generate strong electromagnetic fields (X-ray
or diathermy equipment, MRI machines). Please keep a safe distance of several meters.
The device is not suitable for use in areas with an explosive, flammable or combustive
environment.
During use, the device is to be located in a position allowing direct access to the device“s
central mains supply so that it can be disconnected from the mains at any time.
Inspect the device before use. If there is any damage, it must not be used.
Only accessories provided by Zimmer MedizinSysteme GmbH may be used.
The device may cause malfunctions or may interfere with the operation of equipment in the
vicinity. It may be necessary to take appropriate remedial action, such as realignment, re-
arrangement of the device or install shielding.
During the service life of the device, no changes may be made to the device or medical
system.
The complete medical system is suitable for use in the vicinity of the patient.
To safely disconnect the device from the power supply, pull the mains plug out of the housing
socket or outlet.