Cleaning
Disinfection
11
Page 40
Before starting any maintenance and cleaning measures the device must always be switched
off at the main switch and the mains cable must be disconnected.
Make sure that during cleaning or disinfection no liquid penetrates the device. Do not use
sprays.
If during cleaning or disinfecting liquid penetrates the device, please put the device out of
service, protect it from being used again and contact your service representative.
Make sure that when cleaning and disinfecting the labelling of the device (such as warnings,
labels of control devices, identification plate) is not damaged.
The device and its applied part are considered as uncritical in relation to hygiene due to the
use on non-injured and healthy skin.
Housing / cables
Cleaning: In the event of visible contamination, the housing and all cables can be cleaned
using commercially available alcohol-free plastic cleaners. Wipe the surface until the dirt is
removed, using a soft cloth soaked according to the specifications of the manufacturer of the
cleaning agent but not dripping wet.
Disinfection: We recommend that disinfection is to be carried out at least once a week, as well
as if there is any indication of contamination. Consult with your hygiene specialist when doing
so. Always perform cleaning prior to disinfection.
Housing and all cables can be disinfected using disinfectant wipes. Use a commercially
available alcohol-free disinfectant for metal and plastic, with bactericidal, virucidal and
fungicidal properties. Observe the instructions for use of the respective manufacturer. Wipe all
surfaces using a soft cloth soaked according to the specifications of the manufacturer of the
disinfectant, but not dripping, or with cloths pre-impregnated with disinfectant (wipes).
Also observe requirements for drying or post-cleaning, where applicable.
Transducers
Cleaning: Follow the procedure described under “Housing.
Disinfection: Follow the procedure described under “Housing.