Troubleshooting
Error messages
Disposal
19
Page 52
Cable check
The error message “Cable check” appears on the display.
In the case of electric stimulation therapy in constant current mode, this means:
a break in the patient electrical circuit.
In general, this message refers to electrodes which have fallen off, dirty electrode clips, a
defective patient cable or a defective patient fuse. After eliminating the cause, delete message
with “OK” confirmation.
Overcurrent
The error message “Overcurrent! (Patient protection)” appears on the display.
This refers to an increase in the maximum permitted current. An increase in current during
constant operation generally indicates a device defect, while an increase in current during
constant voltage mode can occur when there is a change in the patient resistance (for
example, wet skin).
Delete message with OK confirmation.
If the error message appears again, please contact Customer Service.
SD card
The error message “Error when opening a file” appears on the display.
This message appears when files stored on the SD card are to be accessed. Insert the card
and confirm with “OK”.
The error message “No SD card found” appears on the display.
This message appears
when programs are to be saved in the Favourites or Memory list and no SD card is
inserted,
when the Favourites or Memory list is to be opened and no SD card is inserted.
Insert the card and confirm with “OK”.
Malfunction
The error message “Error during self-test” appears on the display.
Occasionally, the error can be corrected by switching the device off, waiting 5 seconds, and
switching the device back on. Otherwise please inform Customer Service.
You may get in touch with them via the sales representative or via the main office in Neu-Ulm.
The device may only be returned to the factory in its original packaging.
Main office
Zimmer MedizinSysteme GmbH
Junkersstraße 9
89231 Neu-Ulm, Germany
Tel. +49 731. 9761-291
Fax +49 731. 9761-299
www.zimmer.de
Disposal
The device may only be returned to the factory in its original packaging. It must be disposed
of by the factory in Neu-Ulm.
In foreign (European) countries please refer to national regulations for disposal. Contact your
distributor if necessary.