Service Manual
9650-0309-01 Rev. E
January 2024

Other manuals for ZOLL aed pro

Questions and Answers

  • L
    Lauren QuinnAug 21, 2025
    What to do if my ZOLL Medical Equipment says "CHECK DEFIB PADS"?
    • D
      Douglas RichardsonAug 21, 2025
      If the ZOLL Medical Equipment displays the message "CHECK DEFIB PADS", ensure the defibrillation electrode pads are making proper skin contact and that the patient does not have excessive hair beneath the electrodes. If the message persists, replace the defibrillation cable.
  • A
    Austin HarrisonSep 2, 2025
    What to do if ZOLL aed pro says "RELEASE SHOCK BUTTON"?
    • B
      Betty SimmonsSep 3, 2025
      If your ZOLL Medical Equipment displays the message "RELEASE SHOCK BUTTON", release the Shock button. Wait until the unit issues the prompt PRESS FLASHING SHOCK BUTTON before pressing the button.
  • D
    Dr. Edward SparksSep 1, 2025
    How to fix noisy ECG, artifact, or wandering baseline on ZOLL aed pro Medical Equipment?
    • Y
      yshawSep 1, 2025
      If you are getting a noisy ECG, artifact, or wandering baseline on your ZOLL Medical Equipment, turn off nearby two-way radios and cell phones. Before attaching electrodes, properly prepare the patient’s skin. Check for proper adhesion of the electrodes to the patient. Arrange the cable and leads so that they do not pull on the electrodes or swing excessively.
  • W
    William UnderwoodAug 30, 2025
    Why is my ZOLL aed pro Medical Equipment defibrillator not charging?
    • M
      Mary AndersonAug 31, 2025
      If your ZOLL Medical Equipment is not charging, verify the rhythm. Confirm that the defibrillation cable is plugged in and the pads are attached to the patient. Install a fully charged battery pack.
  • E
    Elizabeth SmithAug 29, 2025
    How to troubleshoot the "CHECK ECG ELECTRODES ATTACH ECG ELECTRODES" message on ZOLL aed pro Medical Equipment?
    • B
      Brittany ThompsonAug 29, 2025
      If your ZOLL Medical Equipment shows the message "CHECK ECG ELECTRODES ATTACH ECG ELECTRODES", ensure that the ECG cable is connected to each electrode and to the unit. Also, make sure that the ECG electrodes are making good contact with the patient and are not dried out. Try replacing the ECG electrodes or the ECG cable.
  • B
    brittanydavisAug 27, 2025
    What does "PLUG IN CABLE" mean on ZOLL aed pro and how to fix it?
    • D
      David SilvaAug 27, 2025
      If the ZOLL Medical Equipment displays the message "PLUG IN CABLE", ensure that the electrode cable is properly connected to the unit. Remove the cable and check for bent or broken pins. Replace the electrode cable if necessary.
  • D
    David KeyAug 26, 2025
    What to do if Ready indicator shows a red “X” on ZOLL Medical Equipment?
    • E
      Evan ColeAug 27, 2025
      If your ZOLL Medical Equipment shows a red “X” while powered on, turn the unit off and then on again. Follow the prompts to resolve the problem. If the unit continues to fail, take it out of service, and contact ZOLL Technical Service.
  • A
    Allison FieldsAug 25, 2025
    What to do if ZOLL Medical Equipment displays "PLUG IN DEFIB CABLE"?
    • C
      chad60Aug 25, 2025
      If the ZOLL Medical Equipment displays the message "PLUG IN DEFIB CABLE", check the cable, and replace if necessary.
  • B
    blake13Aug 23, 2025
    How to fix a "CHANGE BATTERY" message on ZOLL aed pro Medical Equipment?
    • C
      Christopher HarrisAug 23, 2025
      If your ZOLL Medical Equipment displays the message "CHANGE BATTERY," replace the battery pack with a fully charged battery pack as soon as possible.

Summary

Preface

Related Documentation

Describes other manuals available for the AED Pro unit.

Conventions

Explains the manual's formatting, symbols, and warning statements.

Symbols

Lists and describes symbols that may appear in the document or on equipment.

Warranty

Information regarding warranty terms and conditions for the AED Pro.

Safety Considerations

Details essential safety precautions for servicing and operating the AED Pro.

Technical Service

Provides contact information and procedures for technical support.

Returning a Unit for Service

Outlines the process for returning the AED Pro unit for repair.

Chapter 1 Maintenance Tests

1.0 Physical Inspection

Procedure to ensure the unit shows no signs of damage or excessive wear.

2.0 Power Supply Test;Current

Tests the unit's power supply and measures current draw.

3.0 Power ON Test

Verifies the unit powers on correctly and indicates readiness for use.

4.0 Pediatric Mode;Electrode Connector Type Test

Tests pediatric mode functionality and electrode connector type.

5.0 Defibrillator Energy Test

Evaluates the energy output of the defibrillator at various settings.

6.0 Battery Type Test

Verifies the unit's compatibility and function with different battery types.

7.0 Establish Communication

Tests the unit's ability to establish communication with a computer.

8.0 Internal Discharge Test

Performs a test of the unit's internal discharge mechanism.

9.0 Patient Impedance Calibration

Calibrates the unit's patient impedance measurement system.

10.0 ECG Lead-Off Detection

Verifies the unit's capability to detect disconnected ECG leads.

11.0 ECG Offset Calibration

Calibrates the ECG offset level for accurate readings.

12.0 Real Time Clock Test

Tests the accuracy and functionality of the unit's real-time clock.

13.0 Write Calibration to Flash

Saves calibration data to the unit's non-volatile memory.

14.0 Audio Recording Test (if option installed)

Tests the audio recording feature if the option is installed.

15.0 Read Error Log - Final

Retrieves and clears error log messages from the unit.

Chapter 2 Troubleshooting

Troubleshooting the AED Pro Unit

Lists general issues with the unit and their associated corrective actions.

AED Pro Error Log Messages

Lists and explains critical error log messages for the AED Pro unit.

Appendix A Maintenance Tests Checklist

Instructions

Provides guidance on how to use the maintenance tests checklist.

ZOLL AED Pro Maintenance Tests Checklist

The official checklist for recording AED Pro maintenance test results.

ZOLL aed pro Specifications

General IconGeneral
Device TypeAutomated External Defibrillator (AED)
DisplayLCD
Operating Temperature0°C to 50°C (32°F to 122°F)
ECG DisplayYes
WaveformRectilinear Biphasic
Voice PromptsYes
Water ResistanceIP55
IP RatingIP55
ECG MonitoringYes
Warranty5 years
Storage Temperature-20°C to 60°C
Energy SelectionAutomatic, configurable
CPR FeedbackReal CPR Help
Data RecordingYes
Data StorageYes
ConnectivityUSB
Shock EnergySelectable 120J, 150J, 200J
Energy LevelsSelectable 120J, 150J, 200J