c.
Performing an ATP Test run
Follow the Instructions for Use for the
3M™ Clean-Trace™
Tests swabs (UXL100 or AQF100). Collect a sample, activate
the test swab and shake it for at least 5 seconds.
▪ Open the 3M™ Clean-Trace™ Luminometer by pressing the rocker cap. Then insert the activated test into the
chamber. Keep the 3M Clean-Trace Luminometer upright whenever a test is in the chamber to ensure accurate
results and avoid spillage of reagents, which could damage the sensitive components of
the instrument.
▪ Tap [Start] to begin measuring. A progress bar appears and moves from left to right.
▪ The test result appears and is stored in the memory so it can be synchronized with the 3M
Clean-Trace Hygiene Management Desktop Software. The result is also interpreted using
your established Pass/Fail criteria and is displayed on the screen as Pass, Caution, or Fail.
▪ Remove the Test swab from the 3M™ Clean-Trace™ Luminometer; press the rocker cap, grasp the top of the Test swab,
and pull upwards. Follow current industry standards and local regulations for disposal of used Test swabs.
o To test another Test point, tap the next Test point at the bottom right of the LM1 screen, or Previous Test point at
the bottom left of the 3M Clean-Trace LM1 screen.
o To do a retest, tap [Retest] then tap [Start]. Retest is allowed up to 12 hours after the initial test.
o To display the list of Test points, tap [List View].