d.
Performing a Retest on an ATP Test point before Moving to a Different Test point
After completing a test run, the
3M Clean-Trace
LM1 Luminometer offers the option to do a re-test before moving to a
different Test point. To perform a retest, follow the steps below:
▪ Follow the Instructions for Use for the 3M™ Clean-Trace™ Test to collect a sample, activate the test, and shake it for
at least 5 seconds.
▪ Open the 3M Clean-Trace Luminometer by pressing the rocker cap. Then insert the activated test into the chamber.
Keep the 3M Clean-Trace Luminometer upright whenever a test is in the chamber, to ensure accurate results and
avoid spillage of reagents, which could damage the sensitive components of the instrument.
▪ Tap [Retest] then tap [Start] to begin measuring. A progress bar appears and moves from left to right.
▪ The Test result appears and is stored in the memory so it can be synchronized with the 3M Clean-Trace Hygiene
Management Software. The result is also interpreted using established Pass/Fail limits and is displayed on the screen
as Pass, Caution, or Fail.
▪ Remove the test swab from the 3M Clean-Trace Luminometer; press the rocker cap, grasp the top of the Test swab,
and pull upwards. Follow current industry standards and local regulations for disposal of used Test swabs.
e.
Performing a Retest on an ATP Test point after Moving to a Different Test point
The 3M Clean-Trace LM1 Luminometer offers the option to retest a test point within 12 hours from the original test run.
To select the Test point to retest while in the same Sample plan, tap Next [>] or Previous [<] to locate the Test point.
To retest a Test point from a different Sample plan:
▪ Tap (Home), then tap the Sample plan, then tap the date and time of the run.
▪ In the Sample plan, locate then tap the Test point you wish to retest - Scroll down or up as needed.
▪
Follow the Instructions for Use for the 3M Clean-Trace Test to collect a sample, activate the test and shake
it for at least 5 seconds.
▪
Open the 3M Clean-Trace Luminometer by pressing the rocker cap. Then insert the activated test into the chamber.
Keep the 3M™ Clean-Trace™ Luminometer upright whenever a test is in the chamber, to ensure accurate results and
avoid spillage of reagents, which could damage the sensitive components of the instrument.
▪
Tap [Retest] then tap [Start] to begin measuring. A progress bar appears and moves from left to right.
▪
The Test result appears on the screen and is stored in the memory so it can be synchronized with the 3M
Clean-Trace Hygiene Management Software. The result is also interpreted using your established Pass/Fail
limits and is displayed on the screen as Pass, Caution, or Fail.
▪
Remove the Test swab from the 3M Clean-Trace Luminometer; press the rocker cap, grasp the top of the Test swab,
and pull upwards. Follow current industry standards and local regulations for disposal of used Test swabs.