Performing a Retest on an ATP Test point before Moving to a Different Test point
After completing a test run, the
3M Clean-Trace
LM1 Luminometer offers the option to do a re-test before moving to a
different Test point. To perform a retest, follow the steps below:
Follow the Instructions for Use for the 3MClean-Trace Test to collect a sample, activate the test, and shake it for
at least 5 seconds.
Open the 3M Clean-Trace Luminometer by pressing the rocker cap. Then insert the activated test into the chamber.
Keep the 3M Clean-Trace Luminometer upright whenever a test is in the chamber, to ensure accurate results and
avoid spillage of reagents, which could damage the sensitive components of the instrument.
Tap [Retest] then tap [Start] to begin measuring. A progress bar appears and moves from left to right.
The Test result appears and is stored in the memory so it can be synchronized with the 3M Clean-Trace Hygiene
Management Software. The result is also interpreted using established Pass/Fail limits and is displayed on the screen
as Pass, Caution, or Fail.
Remove the test swab from the 3M Clean-Trace Luminometer; press the rocker cap, grasp the top of the Test swab,
and pull upwards. Follow current industry standards and local regulations for disposal of used Test swabs.
e.
Performing a Retest on an ATP Test point after Moving to a Different Test point
The 3M Clean-Trace LM1 Luminometer offers the option to retest a test point within 12 hours from the original test run.
To select the Test point to retest while in the same Sample plan, tap Next [>] or Previous [<] to locate the Test point.
To retest a Test point from a different Sample plan:
Tap (Home), then tap the Sample plan, then tap the date and time of the run.
In the Sample plan, locate then tap the Test point you wish to retest - Scroll down or up as needed.
Follow the Instructions for Use for the 3M Clean-Trace Test to collect a sample, activate the test and shake
it for at least 5 seconds.
Open the 3M Clean-Trace Luminometer by pressing the rocker cap. Then insert the activated test into the chamber.
Keep the 3M Clean-Trace Luminometer upright whenever a test is in the chamber, to ensure accurate results and
avoid spillage of reagents, which could damage the sensitive components of the instrument.
Tap [Retest] then tap [Start] to begin measuring. A progress bar appears and moves from left to right.
The Test result appears on the screen and is stored in the memory so it can be synchronized with the 3M
Clean-Trace Hygiene Management Software. The result is also interpreted using your established Pass/Fail
limits and is displayed on the screen as Pass, Caution, or Fail.
Remove the Test swab from the 3M Clean-Trace Luminometer; press the rocker cap, grasp the top of the Test swab,
and pull upwards. Follow current industry standards and local regulations for disposal of used Test swabs.
Sample plan name
Sample plan run date
Sample plan run time
Results synchronized
Not tested, tap to run test
Tested, PASSED, tap to retest
Total Test Points
(+ Test types)
Tested
Passed
Tested
Caution
Tested
Failed
Tested, FAILED, tap to retest

Table of Contents

Other manuals for 3M Clean-Trace LM1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 3M Clean-Trace LM1 and is the answer not in the manual?

3M Clean-Trace LM1 Specifications

General IconGeneral
Brand3M
ModelClean-Trace LM1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals