EL3000 CONTINUOUS GAS ANALYZERS |  OI/EL3000-EN REV. D  185 
 
Replacing the UV lamp – Installing the new lamp 
1.  Before installing the new lamp, record the serial number 
shown on the name plate 
l. It will need to be provided 
during subsequent configuration and calibration. 
2.  Perform steps 4 to 10. (see Replacing the UV lamp – Remove 
the old lamp on page 183) in reverse order. Observe the 
following information in the process: 
•  Insert the new UV-lamp into the lifted bracket from below 
so that the two guide pins on the bracket fit in the 
appropriate holes 
o in the lamp. 
•  When placing the light protection hood, make sure that 
the connection cables are not crushed or damaged. 
 
Recommissioning the gas analyzer 
1.  Close the housing tightly. 
Notice 
Light penetration during operation leads to erroneous 
measurement values and overranging. 
2.  Turn on the gas analyzer power supply and wait for the 
warm-up phase to end.-up. 
3.  Configure and calibrate the gas analyzer (see Configuring 
and calibrating the gas analyzer on page 185). 
 
 
 
 
 
 
Configuring and calibrating the gas analyzer 
After replacing the UV lamp, the following work must be carried 
out: 
•  Enter the serial number of the new lamp, 
•  Adjust the zero point of the electronic SO
2
 cross-
sensitivity correction if necessary, 
•  Intensity standardization. 
 
Required are: 
•  Software tool Optima TCT Light 
•  Ethernet connection between the gas analyzer and the 
computer 
•  Zero point gas: either nitrogen or sample component-free 
ambient air 
 
Enter new lamp number 
1.  Open the software tool Optima TCT Light. 
2.  Click on the command ‘Configuration mode...’ in the ‘Options’ 
menu. In the ‘Configuration Mode’ dialog, click on ‘AMC 
module’ and close the dialog by clicking ‘OK’. 
3.  Click the command ‘Establish communication...’ in the ‘File’ 
menu. In the ‘Establish communication’ dialog, select the 
‘Network’ communication mode, enter the IP address or the 
server name (see Setting the IP address on page 112) of the 
gas analyzer, click on ‘Connection’ and close the dialog by 
clicking on ‘OK’. In the following query, click ‘Yes’. 
Communication is established and the data from the gas 
analyzer is read.