2101510MNAH | NGC8200 AND PGC1000 USER MANUAL | 77
Figure 7-3: Calibration running
6. When the calibration is complete, the device should move to the designated Next Mode. View
results on the Analyzer Operation screen.
7. The un-normalized total of the calibration stream should be between 99.5% and 100.5%. If
values exceed these parameters, proceed to the Troubleshooting section.
8. Carefully examine the calibration stream’s Chromatogram-1 and Chromatogram-2 by clicking
the button. Look for unlabeled peaks and base line anomalies. Use Figure 7-4 and Figure
7-5 as a guide. If errors exist, proceed to the Troubleshooting section.
9. Once the device is running smoothly, is producing good chromatograms, and all the peaks are
labeled and eluting correctly, perform a save and restore procedure to update tfCold.
Figure 7-4: Typical chromatograph for chromatogram-1 (heavies)