PATIENT TESTING - 6
VerifyNow System User Manual Page 50 14340.J
Figure 6-7 Inspect Humidity Indicator and Prepare Needle
9.
Open the cover. Using the finger grip, insert the assay device into the assay
device port until it clicks (Figure 6-8).
Figure 6-8 Insert Device
10.
If the assay device is the first from a new lot, the instrument will display a bar
code screen as soon as the assay device is inserted into the port and the spot
code is read. (Figure 6-9). Place the assay device pouch in front of the bar code
scanner on the left side of the instrument so that the bar code on the bottom edge
of the pouch lines up with the scanner window. Move the pouch both towards and
away from the instrument. An audible beep will be heard when the instrument
reads the information. Press the Retry key if you are unable to scan the bar
code the first time.