RETRIEVING TEST RESULTS - 7
VerifyNow System User Manual Page 64 14340.J
Feature Description
O1: High optical signal level
O2: Low optical signal level
V1: Maximum vacuum level
V2: Maximum leak rate
P1: Maximum pressure level
F1: Mixer frequency flag count
F2: Mixer frequency transitions high-to-low
Mask: Identifies failing component element
Flagged
Parameter
The failed parameter(s) will be indicated by an arrow () to the
right of the measured value(s). Refer to the Appendix for a
description of parameters.
In addition to viewing earlier results, icons keys enable results to be printed
(optional).
Icon Description
Print Result Scroll to the selected record, and select this icon to print the
selected EQC results.
Print All
Results
Scroll backwards through the results using the
key, and
select this icon to print all EQC results in the log that are
forward of the cursor position.
Back Select this icon to return to the previous screen.
7.2.2 Wet Quality Control Results Log
The cursor is positioned on the most recent assay result (Figure 7-5). Navigate up
and down an individual list using the (
,