This chart lists the causes of the error messages and what to do to solve them:
Code Message Description Resettable
E101 memory fail EEPROM fault NO
E102 memory fail I2C bus EEPROM fault NO
E103 memory fail EEPROM DMA readings fault NO
E104 memory fail EEPROM DMA writings fault NO
E105 no response from generator
Lost communication between inverter and
generator
NO
E106 wrong response from generator NAK response from inverter YES
E107 emission start fail YES
E108 emission too long YES
E109 button released beforehand YES
E110 wrong use of button
Exposure button pressed when “ready” mode
not active.
YES
E111 keyboard pressed at boot Keyboard pressed during boot time NO
E112 button pushed at boot Exposure button pressed beforehand NO
E113 wrong use of keyboard Keys pressed during an exposure YES
E114 button release timeout
Exposure button released beforehand after an
exposure
YES
E115 wrong parameters in generator
Exposure parameters (kV, mA, exposure time,
mode) does not match with the ones set in the
control box
YES
E116 generator reset during emission inverter has been resetted during an exposure YES
E117 calibration start fail Calibration YES
E118 calibration too long Calibration time too long YES
E119 button pushed during cooling
Exposure button pressed beforehand pushed
during a functionality exposure
YES
E120 cooling time disabled Functionality pause disabled YES
E121 button pushed during stand-by
Exposure button pressed beforehand during
stand-by
YES
E122 memory fail Parameters out of range NO
E123 power relay fail power relay failure (relay OFF) NO
E124 power relay fail power relay failure (relay ON) NO
E125 power board fail power board failure NO
E126 power board fail power board failure NO
E127 power board fail power board failure NO
E128 power board fail power board failure NO
E129 wireless unit not connected wireless receiver board not connected YES
E132 system error hardware error NO