E201 generator memory Inverter EEPROM error NO
E202 generator memory Inverter I2C EEPROM error NO
Code Message Description Peut être réinitialisé
E203 generator memory Inverter EEPROM DMA readings fault NO
E204 generator memory Inverter EEPROM DMA writings fault NO
E205 generator not calibrated Calibration not yet done YES
E206 tube voltage too low Anodic voltage too low YES
E207 tube voltage too high Anodic voltage too high YES
E208 electrical discharge on hv YES
E209 tube current too low Anodic current too low YES
E210 tube current too high Anodic current too high YES
E211 lament voltage too low Filament voltage too low YES
E212 lament voltage too high Filament voltage too high YES
E213 tube unit not connected High voltage unit not connected NO
E214 button released beforehand
Enable signal released beforehand during an
exposure
YES
E215 generator internal power supply 12V internal voltage too low NO
E216 generator internal power supply 12V internal voltage too high NO
E217 generator mode Wrong exposure enabling signal activation YES
E218 safe circuit fail Security interlock fault NO
E219 tube current too low Mean anodic current too low YES
E220 tube current too high Mean anodic current too high YES
E221 no tube voltage feedback Missing HV feedback NO
E222 ACE acquisition fail X-rays not detected by Sopix YES
E223 tube unit temperature too high Monoblock over temperature YES
E224 tube unit temperature too low Monoblock under temperature YES
E225 tube unit temp. sensor fail
Monoblock temperature sensor fault (open
circuit)
YES
E226 tube unit temp. sensor fail
Monoblock temperature sensor fault (short
circuit)
YES
E227 voltage reference fail generator voltage reference too low YES
E228 voltage reference fail generator voltage reference too high YES
E232 system error hardware error NO
For the errors that can’t be cleared by the reset button on the keypad, please contact your installer or your local Acteon customer service