8230 Optical Power Meter Operation Manual
4.6.2 Remote Control Command List
4-18
Remote Header H0 Sets the header output to OFF
H1 Sets the header output to ON
H? Query reply : S0 to S1
Block delimiter DL0 Sets the block delimiter to CR+LF.
DL1 Sets the block delimiter to LF.
*1
DL? Query reply : DL0 to DL1
Status *STB? Reply to the status byte register query: ddd
*SRE nnn Sets the service request enable register. (0 to 255) ( 0 ) ( 0 )*1
*SRE? Query reply : ddd
*ESR? Reply to the standard event status register query:
ddd
*ESE nnn Sets the standard event status enable register. (0 to
255)
( 0 ) ( 0 )*1
*ESE? Query reply : ddd
DSR? Reply to the device event status register query:
ddddd
DSE nnnnn Sets the device event status enable register. (0 to
65535)
( 0 ) ( 0 )*1
DSE? Query reply : ddddd
ERR? Reply to the error register query: ddddd
*CLS Clears the status
System Initialization C Resets the system to the conditions when the
power was turned on. (Set parameters do not
change.) *2
*RST Sets the system conditions when the power was
turned on and the set parameters to the factory
default values. *2*3
Changing the
response to
*IDN?
*4
OID0 The new model name is the response to *IDN?.
OID1 The old model name is the response to *IDN?.
OID? Response: OID0 or OID1
*1: Cannot be initialized by the RL command.
*2: The data, which is saved in the output buffer, is deleted.
*3: MYID and the key lock status are not changed.
*4: This command is applied to software revision D00 or later.
Table 4-5 Remote Control Commands (4 of 7)
Item Command Description
Initial value Availability
Power
ON
Factory
default
Calibration
mode
Error2