EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>1260 Infinity II Refractive Index Detector

Agilent Technologies 1260 Infinity II Refractive Index Detector User Manual

Agilent Technologies 1260 Infinity II Refractive Index Detector
217 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #111 background imageLoading...
Page #111 background image
Agilent InfinityLab LC Series RID User Manual 111
8 Test Functions
Refractive Index Calibration
6
Fill the sample cell with calibration solution.
a
Remove the inlet capillary or flushing syringe from the in port.
b
Take the syringe and fix the needle to the syringe adapter.
c
Suck about 1.5 mL of the calibration sample into the syringe.
d
Keep the syringe in a horizontal position.
e
Remove the needle.
f
Add the filter to the syringe and fit the needle to filter.
Figure 20 Syringe with Sample Filter
g
Lift the needle tip and carefully eject approximately 0.5 mL to remove air
out of the syringe and to flush the needle.
h
Add the PEEK fitting to the needle tip and fix both at the flow cell inlet.
i
Slowly inject about 1.0 ml and wait for about 10 s to inject another 0.1 mL.
This will assure that the cell is filled properly.
7
Leave the RID Tools Screen.
Sample filter
NOTE
Do not inject the calibration solution without the sample filter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals