Agilent InfinityLab LC Series RID User Manual 187
13 LAN Configuration
Initialization Mode Selection
Using Default
When
Using Default
is selected, the factory default parameters are taken instead.
These parameters enable a TCP/IP connection to the LAN interface without
further configuration, see Table 18 on page 187.
Figure 35 Using Default (Principle)
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the module must reside in the same
subnet.
The user may open a Telnet session using the default IP address and change the
parameters stored in the non-volatile memory of the module. He may then close
the session, select the initialization mode Using Stored, power-on again and
establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a
local hub), separated from the local area network, the user may simply keep the
default parameters to establish the TCP/IP connection.
Active
Parameter
Default
Parameter
Using the default address in your local area network may result in network
problems. Take care and change it to a valid address immediately.
Table 18 Using Default Parameters
IP address: 192.168.254.11
Subnet Mask: 255.255.255.0
Default Gateway not specified
In the
Using Default
mode, the parameters stored in the memory of the module
are not cleared automatically. If not changed by the user, they are still available,
when switching back to the mode Using Stored.