12 16901A Logic Analysis System Service Guide
1 General Information
Electrical and Operating Environment Characteristics
The following operating characteristics are not specifications,
but are typical operating characteristics for the Agilent
16901A logic analysis system.
Electrical Characteristics
Power Requirements Frame (16901A): 115/230 Vac +/- 20%, 48-66Hz, 775W Max
The line voltage is autodetected by the instrument.
CAT II (Line voltage in appliance and to wall outlet)
Pollution degree 2
Environmental Characteristics
For indoor use only.
Tem per at ure Instrument (16901A): 0° to + 45° C (+32° to +113° F)
Probes/cables: 0° to + 65° C (+32° to +149° F)
Altitude 3,000 m (10,000 ft)
Humidity Relative humidity 8% to 80% at 40° C (104° F). Avoid sudden, extreme temperature changes which
could cause condensation on the circuit board.
Vibration Operating: random vibration 0-500 Hz, 10 minutes per axis, 0.3 g (rms).
Non-operating: random vibration 0-500 Hz, 10 minutes per axis, 2.41g (rms); and swept sine resonant
search, 0-500 Hz, 0.75g, 5-minute resonant dwell at 4 resonances per axis.
Weight Model Max Net Max Shipping
16901A 13.6 kg (30.0 lbs) 20.4 kg (45.0 lbs)
The weight of modules ordered with mainframes will add 0.9 kg (2.0 lb) per module.
Storage Store or ship the instrument in environments within the following limits:
Temperature -40°C to +75°C (-40°F to 167°F)
Humidity up to 90% at 65° C (149°F)
Altitude up to 3,000 meters (10,000 feet)
Protect the system from temperature extremes which cause condensation on the instrument.