Replacing Assemblies 6
16901A Logic Analysis System Service Guide 51
To remove and replace the power supplies
600 watt power supply
The 600 watt power supply is on all models but may look
different than the one shown here.
1 Perform previous procedures:
“To prepare the instrument for disassembly” on
page 46.
“To remove and replace the cover” on page 50.
2 Press the release latches on the sides of the connector to
disconnect the top 600 watt power supply cable from the
module interface board (MIB).
3 Press the single release tab to disconnect the lower 600
watt power supply cable from the MIB.
4 Press the release latches to disconnect the 600 watt line
filter cable.
WARNING
Hazardous voltages exist on the power supply. To avoid electrical
shock, disconnect the power from the instrument before
performing the following procedures. After disconnecting the
power, wait at least six minutes for the capacitors on the power
supply board to discharge before servicing the instrument.

Table of Contents

Other manuals for Agilent Technologies 16901A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 16901A and is the answer not in the manual?

Agilent Technologies 16901A Specifications

General IconGeneral
BrandAgilent Technologies
Model16901A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals