Testing Performance 3
16901A Logic Analysis System Service Guide 21
To perform the power-up tests
The logic analysis system automatically performs power- up
tests when you apply power to the instrument. Any errors
are reported in the boot dialogue. Serious errors will
interrupt the boot process.
The power- up tests are designed to complement the
instrument on- line Self- Tests. Tests that are performed
during powerup are not repeated in the Self- Tests.
A keyboard and mouse must be connected to the mainframe
to observe the results of the power- up tests.
1 Exit all logic analysis sessions.
2 Press the power switch for about 1 second and release,
wait for the unit to power off.
3 After a few seconds, turn the power switch back on.
Observe the boot dialogue for the following:
Ensure all of the installed memory is recognized
Any error messages
Interrupt of the boot process with or without error
message
A complete transcript of the boot dialogue is in chapter 8,
"Theory of Operation."
4 During initialization, check for any failures.
If an error or an interrupt occurs, refer to Chapter 5,
“Troubleshooting.”

Table of Contents

Other manuals for Agilent Technologies 16901A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 16901A and is the answer not in the manual?

Agilent Technologies 16901A Specifications

General IconGeneral
BrandAgilent Technologies
Model16901A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals