Chapter 4 Calibration and Adjustment
Self-Test
4
4
Self-Test
Self-Test is performed as the first step to ensure the instrument is in
working order before beginning any additional adjustments.
Note Be sure to follow the requirements listed in “Test Considerations” on
page 91 before beginning any adjustments.
1
Press
and then press the
Calibrate
softkey. Enter setup number
“1” and select BEGIN.
2 If the instrument fails any self-test, you must repair the instrument
before continuing the adjustment procedures.
Note The self-test procedure takes approximately 15 seconds to complete.
Setup
1 Performs the Self-test. The Main Output is disabled during test.